Factors Affecting the Primary Stability of the Intraosseous Screw Type Implants /// Фактори повлияващи първичната стабилност на интраосалните винтови имплантати
Date
2016Author
Sabeva, Elitsa
Събева, Елица
elitsa.sabeva@mu-varna.bg
Metadata
Show full item recordAbstract
[EN] The aim of this study was to establish the influence of the implant design, diameter, length, thread pitch and surface modification on the primary stability and on the heat generation at the marginal bone area during the implant insertion. Assessment of the primary stability of implants of different forms and dimensions inserted into artificial bone blocks, into pig ribs and into 3D-printed simulation models of lower jaw, was performed. Stability was assessed by the following methods: measurement of maximum insertion torque (MIT), damping capacity and resonance frequency analysis, and heat generation during the insertion of implants recorded by a thermal camera. [BG] Целта на дисертационния труд е да се установи влиянието на формата, диаметъра, дължината, повърхностната обработка и дизайна на резбата на интраосалните винтови имплантати върху първичната им стабилност и генерирането на топлина в областта на кресталната кост по време на тяхното поставяне. За постигане на тази цел се измери първичната стабилност на имплантати с различни размери и форма, поставени в блокчета изкуствена кост, свински ребра и 3D-принтирани симулационни модели на долна челюст. Стабилността се оценява чрез следните методи: измерване на максималния въртящ момент при поставяне, амортизиращия капацитет и резонансно-честотен анализ, а генерираната топлина по време на поставянето се регистрира чрез термокамера.